406510 Neogen Prostaglandin E2 (Monoclonal)

  • Email link to this assay page
Neogen

This data analysis tool calculates sample concentrations using a Four Parameter Logistic (4PL) curve fit in accordance with Neogen’s Prostaglandin E2 (Monoclonal) (406510) kit.

This analysis optionally includes a substrate background correction step. If a blank group is defined on your layout, the mean of the blank replicates is first subtracted from the raw data measurements.

The maximal binding (%B/B0) is calculated for each sample using the B0 measurement.

The standard data points (concentration vs. %B/B0) are plotted on semi-log axes and a 4PL curve is fitted through the points. The concentrations of the samples are determined from the fit with any specified dilution factors applied.

The %CV, Standard Deviation and Standard Error are calculated for each replicated sample.

Samples outside the range of the standards or the fit (greater than the upper asymptote or below the lower asymptote) are highlighted in yellow.